IEC 61010-031 Ed. 2.0 Edition 05/2015
Safety requirements for electrical equipment for measurement, control and laboratory use - Part 031: Safety requirements for hand-held probe assemblies for electrical measurement and test
  •   
  •  
  • 345.6 / Kopie
  •  
 

Abstract

IEC 61010-031:2015 specifies safety requirements for hand-held and hand-manipulated probe assemblies of the types described below, and their related accessories. These probe assemblies are for direct electrical connection between a part and electrical test and measurement equipment. They may be fixed to the equipment or be detachable accessories for the equipment. It has the status of a group safety publication in accordance with IEC GUIDE 104. IEC 61010-031 is a stand-alone standard. This second edition cancels and replaces the first edition published in 2002 and Amendment 1:2008. This edition constitutes a technical revision. This edition includes the following significant changes from the first edition, as well as numerous other changes:
a) Voltages above the levels of 30 V r.m.s., 42,4 V peak, or 60 V d.c. are deemed to be HAZARDOUS LIVE instead of 33 V r.m.s., 46,7 V peak, or 70 V d.c.
b) Servicing is now included within the scope.
c) Extended environmental conditions are included within the scope.
d) New terms have been defined.
e) Tests for REASONABLY FORESEEABLE MISUSE have been added, in particular for fuses.
f) Additional instruction requirements for probe assembly operation have been specified.
g) Limit values for ACCESSIBLE parts and for measurement of voltage and touch current have been modified.
h) SPACINGS requirements for mating of CONNECTORS have been modified.
i) PROBE TIPS and SPRING-LOADED CLIPS requirements have been modified. The PROTECTIVE FINGERGUARD replace the BARRIER with new requirements.
j) Insulation requirements (6.5) and test procedures (6.6.5) have been rewritten and aligned when relevant with Part 1. Specific requirements have been added for solid insulation and thin-film insulation.
k) The terminology for MEASUREMENT CATEGORY I has been replaced with the designation "not RATED for measurements within MEASUREMENT CATEGORIES II, III, or IV".
l) The flexing/pull test (6.7.4.3) has been partially rewritten.
m) Surface temperature limits (Clause 10) have been modified to conform to the limits of IEC Guide 117.
n) Requirements for resistance of PROBE WIRES to mechanical stresses have been added in Clause 12 and a new Annex D.
o) Requirements have been added regarding the prevention of HAZARD from arc flash and short-circuits for SPRING-LOADED CLIPS.
p) A new informative Annex E defines the dimension of the 4 mm banana CONNECTORS.

Status

Standard - Aktiv

Ursprung

Technisches Komitee :
66 : Safety of measuring, control and laboratory equipment

Annahme

Beginn der Abstimmung über den Entwurf      Datum der Ratifizierung (dor)   
Ende der Abstimmung über den Entwurf      Datum der Ankündigung (doa)   
Beginn der Abstimmung über den Schlussentwurf      Datum der Veröffentlichung (dop)   
Ende der Abstimmung über den Schlussentwurf      Datum der Zurückziehung (dow)   


Veröffentlichung im Amtsblatt
des Grossherzogtum Luxemburg
Referenz

Relations

Evolutions
IEC 60951-3 Ed.3.0 RLV
IEC 61010-031 Ed.3.0

Internationale Normungsklassifizierung (ICS) :

19.080 : Electrical and electronic testing